FDA-Device2014-05-21Class I
Alaris Pump model 8100 with software version 9.1.18 and software upgrade kits with "Delay Until" Option and "Multidose" Feature The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural.
CareFusion 303, Inc.
Hazard
CareFusion is recalling the Alaris Pump model 8100 version 9.1.18 because it may have a software issue that results in situation where the pump module will not properly delay an infusion when the "Delay Until" option or "Multidose" feature is used.
Sold states
Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AU, and KW.
Affected count
57,219 units total (55,756 units in US)
Manufactured in
10020 Pacific Mesa Blvd, San Diego, CA, United States
Products
Alaris Pump model 8100 with software version 9.1.18 and software upgrade kits with "Delay Until" Option and "Multidose" Feature The Alaris Pump module is a large volume infusion pump offered under the Alaris System. The Alaris Pump module will deliver medication and fluids using IV administration sets for continuous or intermittent delivery through clinically acceptable routes of administration such as intravenous or epidural.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1567-2014Don't want to check this manually?