FDA-Device2025-04-23Class II
Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).
Maquet Cardiovascular, LLC
Hazard
Incorrect resistor utilized in the VH-3010 Power Supply, which may cause the jaws of the HemoPro 2 to not adequately heat up.
Sold states
Worldwide - US Nationwide distribution in the states of AL, CA, CO, DC, FL, GA, IA, ID, IL, IN, KY, MD, MI, MN, MO, NC, NH, NJ, NY, OH, PA, PR, TX, UT, VA, WI, WV and the countries of China, Japan, Netherlands, New Zealand, Singapore, United Kingdom, United States.
Affected count
110 units
Manufactured in
45 Barbour Pond Dr, Wayne, NJ, United States
Products
Maquet VH-3010 Power Supply. Used to deliver power to the harvesting tool of the Vasoview Hemopro EVH System (HemoPro 2).
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1577-2025Don't want to check this manually?