Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426
Randox Laboratories Ltd.
Hazard
Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.
Sold states
US Nationwide distribution in the states of GA, MA, WV including PR.
Affected count
6 units
Manufactured in
55 Diamond Road, Crumlin (North), Ireland
Products
Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426
Official notice