FDA-Device2024-04-24Class II

Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426

Randox Laboratories Ltd.
Hazard

Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.

Sold states
US Nationwide distribution in the states of GA, MA, WV including PR.
Affected count
6 units
Manufactured in
55 Diamond Road, Crumlin (North), Ireland
Products
Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1594-2024

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief