FDA-Device2019-06-05Class II

Maquet Getinge-BO-T 2303 Small Adult Cardiac Surgery Material:709000004

Maquet Cardiovascular, LLC
Hazard

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

Sold states
Nationwide
Affected count
N/A
Manufactured in
45 Barbour Pond Dr, N/A, Wayne, NJ, United States
Products
Maquet Getinge-BO-T 2303 Small Adult Cardiac Surgery Material:709000004

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1595-2019