FDA-Device2024-04-24Class II

SW Bari Lift & Transfer, Model Number 38060000

Raye's Inc.
Hazard

Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.

Sold states
Worldwide distribution - US Nationwide and the countries of Australia, UK.
Affected count
1,149 units
Manufactured in
500 Commerce Pkwy, Hays, KS, United States
Products
SW Bari Lift & Transfer, Model Number 38060000

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1595-2024