FDA-Device2016-05-11Class II
MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.
Siemens Medical Solutions USA, Inc
Hazard
The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.
Sold states
US nationwide distribution to OH, FL, TN, MN, OH, NY, VA, and MN.
Affected count
8 systems
Manufactured in
40 Liberty Blvd, Malvern, PA, United States
Products
MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.