FDA-Device2021-05-26Class II
Medtronic REF: MMT-7008B, Enlite Glucose Sensor for the following packaging: ENLITE PLUS 23L 1CL / GTIN: 00643169541719 ENLITE PLUS 2L 1CL PR/ GTIN: 00643169786738 ENLITE PLUS 1PK 23L 1CL/ GTIN: 00763000421304 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.
Medtronic Inc.
Hazard
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Sold states
Worldwide distribution.
Affected count
20,913 units
Manufactured in
18000 Devonshire St, N/A, Northridge, CA, United States
Products
Medtronic REF: MMT-7008B, Enlite Glucose Sensor for the following packaging: ENLITE PLUS 23L 1CL / GTIN: 00643169541719 ENLITE PLUS 2L 1CL PR/ GTIN: 00643169786738 ENLITE PLUS 1PK 23L 1CL/ GTIN: 00763000421304 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.