FDA-Device2021-05-26Class II
Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000337049 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.
Medtronic Inc.
Hazard
Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Sold states
Worldwide distribution.
Affected count
302,667 units
Manufactured in
18000 Devonshire St, N/A, Northridge, CA, United States
Products
Medtronic REF: MMT-7020B, Guardian Sensor (3) for the following packaging: GS3 1PK 2L US/GTIN: 00763000140779 GS3 1PK 2L US/ GTIN: 00763000337049 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.
Official notice