FDA-Device2016-05-11Class II

Alaris PC unit, Model 8015 The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.

CareFusion 303, Inc.
Hazard

The Alaris PC units model 8015 may display a system error code 133.6080 due to failure with the super capacitor (C245) at power up on the Alaris PC unit logic boards.

Sold states
Worldwide Distribution: US (nationwide) and countries of: Australia, Great Britain and Canada.
Affected count
23,397 units
Manufactured in
10020 Pacific Mesa Blvd, San Diego, CA, United States
Products
Alaris PC unit, Model 8015 The Alaris PC unit is the central programming, monitoring and power supply component for the Alaris System.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1606-2016