FDA-Device2021-05-26Class II

Medtronic REF: MMT-7020D3 Guardian Sensor (3) for the following packaging: GS3 1 PK OUS3/ GTIN: 00643169704121 GS3 1PK 9L OUS3/ GTIN: 00763000318895 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.

Medtronic Inc.
Hazard

Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Sold states
Worldwide distribution.
Affected count
13,014 units
Manufactured in
18000 Devonshire St, N/A, Northridge, CA, United States
Products
Medtronic REF: MMT-7020D3 Guardian Sensor (3) for the following packaging: GS3 1 PK OUS3/ GTIN: 00643169704121 GS3 1PK 9L OUS3/ GTIN: 00763000318895 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.