FDA-Device2013-07-10Class II
StatLock CV Plus w/Pigtail, Catheter Stabilization Device Sterile EO Single Use Only, C.R. Bard, Inc. The StatLock¿ device provides stabilization for compatible medical tubes and catheters.
Bard Access Systems
Hazard
BARD Access Systems is recalling lot JUWJF283 of StatLock CV Plus Stabilization Butterfly Devices due to incorrect pad assembly used during manufacture. The absence of the hydrocolloid strip on the bottom may pose an incremental risk including localized skin irritation. Missing directional arrow may lead to mis-orientation of the StatLock stabilization which may impact the securement of the cathet
Sold states
National and Global distribution (Europe, European Economic Area, Argentina, China, and Brazil).
Affected count
3,875 units.
Manufactured in
605 North 5600 West, N/A, Salt Lake City, UT, United States
Products
StatLock CV Plus w/Pigtail, Catheter Stabilization Device Sterile EO Single Use Only, C.R. Bard, Inc. The StatLock¿ device provides stabilization for compatible medical tubes and catheters.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1618-2013Don't want to check this manually?