FDA-Device2020-04-08Class II
Philips V60 Ventilators with Power Management PCBA part number 1055906
Respironics California, LLC
Hazard
Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia
Sold states
U.S.: CO, PA, AZ, OH, MT, AR, FL, NC, KS, IL, CA, HI, MD, SC, WY, OR, NY, NE, OK, MN, UT, ND, LA, IN, AL, AK, WV, GA, WI, MI, VA, NJ, TX, WA, SD, KY, TN, MS, MO, DE, MA, VT, CT, IA, NV, ID, ME, NM, DC, NH, VI, RI, PR. O.U.S.: Argentina, Australia, Austria, Bahrain, Belgium, Bolivia, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Gibraltar, Hong Kong, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Jersey, Kazakhstan, Kenya, Korea, Republic of, Kosovo, Kuwait, Lebanon, Libya, Luxembourg, Malaysia, Maldives, Malta, Mexico, Morocco, Namibia, Netherlands, New Zealand, Nigeria, Norway, Oman, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Russian Federation, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, Viet Nam
Affected count
23581
Manufactured in
2271 Cosmos Ct, N/A, Carlsbad, CA, United States
Products
Philips V60 Ventilators with Power Management PCBA part number 1055906
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1622-2020Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief