FDA-Device2020-04-15Class II

Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)

Zeiss, Carl Inc
Hazard

When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system

Sold states
Worldwide Distribution - US Nationwide
Affected count
US: 600; Global: 1600
Manufactured in
1 Zeiss Dr, N/A, Thornwood, NY, United States
Products
Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1629-2020