FDA-Device2023-06-14Class I

StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systems

Medtronic Navigation, Inc.
Hazard

During non-axial/some axial exams, software anomalies occur during procedures affecting depth gauge graphic displays for cranial biopsy; causing displays to no longer synchronize with other navigational information showing inaccurate values, which may result in prolonged/additional procedure, tissue injury

Sold states
Worldwide Distibution: US (nationwide): OH, IL, CA, MI, FL, LA, OK, WA, AR, NJ, MN, MO, WI, MA, ND, NY, MD, OR, AK, NE, TN, AZ, IN, GA, GU, VA, PA, TX, SD, MT, NV, MS, KY, AL, CT, WV, KS, IA, NC, UT, CO, NH, DC, WY. OUS (foreign): ALBANIA, ARGENTINA, Australia, AUSTRIA, BANGLADESH, BARBADOS, BELGIUM, Bosnia and Herzegovina, BULGARIA, CANADA, CHILE, CROATIA, CZECH REPUBLIC, DENMARK, FRANCE, GREECE, INDIA, IRELAND, ITALY, JAPAN, LUXEMBOURG, MARTINIQUE, NEW ZEALAND, POLAND, PORTUGAL, Republic of Korea, ROMANIA, RUSSIAN FEDERATION, SINGAPORE, SLOVAKIA, SPAIN, SWITZERLAND, TAIWAN, Turkey, UNITED ARAB EMIRATES, UNITED KINGDOM
Affected count
746
Manufactured in
200 Medtronic Drive, Lafayette, CO, United States
Products
StealthStation Cranial Software, Models: 9735585, 9735586 (kit), 9735587 (kit), used with StealthStation S7/i7 systems

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1630-2023

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief