FDA-Device2015-05-27Class II

EPIQ 5 Ultrasound System versions 1.3.2 or lower, WITHOUT the Pediatric Cardiology option, Model: EPIQ 5G, EPIC 5C, EPIQ 5W.Catalog Number: 795204 / 795205. Part Number: 989605408541 With 453561736761 (1.1) 453561750041 (1.1.1) 453561753651 (1.1.2) 453561772231 (1.2) 453561772611 (1.2.1) 453561786571 (1.2.2) 453561785081 (1.3) 453561800581 (1.3.1) 453561805181 (1.3.2) Diagnostic ultrasound imaging and fluid flow analysis.

Philips Ultrasound, Inc.
Hazard

When EPIQ 5 Ultrasound System, WITHOUT the Pediatric Cardiology option, is set up to "Metric" and weight and/or height is entered, a unit conversion error may result in the incorrect display of their values and incorrect calculation of the patients Body Surface Area (BSA).

Sold states
Worldwide Distribution
Affected count
Total 1108 units (490 units in US and 618 units outside the US)
Manufactured in
22100 Bothell Everett Hwy, N/A, Bothell, WA, United States
Products
EPIQ 5 Ultrasound System versions 1.3.2 or lower, WITHOUT the Pediatric Cardiology option, Model: EPIQ 5G, EPIC 5C, EPIQ 5W.Catalog Number: 795204 / 795205. Part Number: 989605408541 With 453561736761 (1.1) 453561750041 (1.1.1) 453561753651 (1.1.2) 453561772231 (1.2) 453561772611 (1.2.1) 453561786571 (1.2.2) 453561785081 (1.3) 453561800581 (1.3.1) 453561805181 (1.3.2) Diagnostic ultrasound imaging and fluid flow analysis.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1631-2015

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.