FDA-Device2015-05-27Class II

EPIQ 7 Ultrasound System versions 1.3.2 or lower, WITHOUT the Pediatric Cardiology option, Model: EPIQ 7G, EPIC 7C, EPIQ 7W. Catalog Number: 795200 / 795201 Part Number: 989605386721 With: 453561726491 (1.0) 453561728121 (1.0.1) 453561736781 (1.1) 453561750021 (1.1.1) 453561753631 (1.1.2) 453561772251 (1.2) 453561772631 (1.2.1) 453561786591 (1.2.2) 453561785101 (1.3) 453561800601 (1.3.1) 453561805211 (1.3.2) Diagnostic ultrasound imaging and fluid flow analysis

Philips Ultrasound, Inc.
Hazard

When EPIQ 7 Ultrasound System, WITHOUT the Pediatric Cardiology option, is set up to "Metric" and weight and/or height is entered, a unit conversion error may result in the incorrect display of their values and incorrect calculation of the patients Body Surface Area (BSA).

Sold states
Worldwide Distribution
Affected count
1692 units total (902 units in the US and 790 units outside the US)
Manufactured in
22100 Bothell Everett Hwy, N/A, Bothell, WA, United States
Products
EPIQ 7 Ultrasound System versions 1.3.2 or lower, WITHOUT the Pediatric Cardiology option, Model: EPIQ 7G, EPIC 7C, EPIQ 7W. Catalog Number: 795200 / 795201 Part Number: 989605386721 With: 453561726491 (1.0) 453561728121 (1.0.1) 453561736781 (1.1) 453561750021 (1.1.1) 453561753631 (1.1.2) 453561772251 (1.2) 453561772631 (1.2.1) 453561786591 (1.2.2) 453561785101 (1.3) 453561800601 (1.3.1) 453561805211 (1.3.2) Diagnostic ultrasound imaging and fluid flow analysis

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1632-2015