FDA-Device2014-05-28Class II

GE Healthcare, NeuroMuscular Transmission Module, E-NMT-01 Product Usage: The Datex-Ohmeda S/5TM E-NMT module is indicated for monitoring the relaxation of the patient and regional block stimulation for nerve location.

GE Healthcare, LLC
Hazard

When the E-NMT-01 module is used in conjunction with the ElectroSensor, the Neuromuscular Transmission (NMT) values may indicate a deeper level of muscle relaxation than the actual level of muscle relaxation. In the clinical situation visual movements of the hand are seen after TOF (Train of Four) stimulation, but the patient monitor shows no counts, or counts are not corresponding to the actual a

Sold states
US Nationwide Distribution in the states of: AR, CA, CO,DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MI, MN, MO, NC, NV, NY, OH, OK, TN, TX, UT, WA, WI, WY. OUS: ARGENTINA. AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, CANADA, CHILE, COLOMBIA, COSTA RICA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, HONG KONG, HUNGRY, INDIA, IRA, ISRAEL, ITALY, JAPAN, KAZAKTSTAN, KUWAIT, LATVIA, LITHUANIA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PANAMA, POLAND, QATAR, ROMANIA, RUSSIAN, F¿D¿RATION, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SPAIN, SWEDEN, SWITZERLAND, TURKEY, TAIWAN, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VENEZUELA.
Affected count
1,794 (471 units US, 1,323 units OUS)
Manufactured in
3000 N Grandview Blvd, N/A, Waukesha, WI, United States
Products
GE Healthcare, NeuroMuscular Transmission Module, E-NMT-01 Product Usage: The Datex-Ohmeda S/5TM E-NMT module is indicated for monitoring the relaxation of the patient and regional block stimulation for nerve location.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1640-2014