FDA-Device2023-06-14Class I
Trilogy Evo, Garbin Evo, and Aeris EVO, Model Numbers DS2110X11B, IN2110X15B, JP2110X16B, AU2110X15B, LA2110X15B, CA2110X12B, BR2110X18B, KR2110X15B, IA2110X15B, FP2110X10, FR2110X14B, ND2110X15B, IT2110X21B, ES2110X15B, DE2110X13B, BL2110X15B, GB2110X15B, EU2110X15B, EE2110X15B, LD2110X23B, and VT2110X24B
Philips Respironics, Inc.
Hazard
Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.
Sold states
US Nationwide. Global Distribution.
Affected count
57,395 devices
Manufactured in
1001 Murry Ridge Ln, Murrysville, PA, United States
Products
Trilogy Evo, Garbin Evo, and Aeris EVO, Model Numbers DS2110X11B, IN2110X15B, JP2110X16B, AU2110X15B, LA2110X15B, CA2110X12B, BR2110X18B, KR2110X15B, IA2110X15B, FP2110X10, FR2110X14B, ND2110X15B, IT2110X21B, ES2110X15B, DE2110X13B, BL2110X15B, GB2110X15B, EU2110X15B, EE2110X15B, LD2110X23B, and VT2110X24B