FDA-Device2014-06-04Class II

ADVIA Centaur Systems VitD 500 test, Ready Pack, Catalog Number: 10631021. Intended for the invitro diagnostic use in quantitative determination of various assays on the ADVIA Centaur and ADVIA Centaur XP analyzers.

Siemens Healthcare Diagnostics, Inc
Hazard

Samples containing fluorescein may show interference with the ADVIA Centaur¿ Systems TSH3 Ultra, Vitamin D and the ADVIA Centaur BRAHMS Procalcitonin assays.

Sold states
Worldwide Distribution-USA (nationwide) and the countries of Afghanistan Algeria Armenia Austria Bahrain Belgium Bulgaria Burkina Faso Canary Islands Croatia Cyprus Czech Republic Denmark Estonia Finland France Germany Great Britain Greece Hungary Iceland Iran Iraq Ireland Israel Italy Jordan Kazakhstan Latvia Lebanon Lithuania Luxembourg Macedonia Malta Netherlands Norway Oman Pakistan Poland Portugal Qatar Romania Russian Fed. Saudi Arabia Serbia Slovakia Slovenia Spain Sweden Switzerland Turkey Unit.Arab Emir. Vatican Yemen.
Affected count
12,787 kits
Manufactured in
333 Coney St, East Walpole, MA, United States
Products
ADVIA Centaur Systems VitD 500 test, Ready Pack, Catalog Number: 10631021. Intended for the invitro diagnostic use in quantitative determination of various assays on the ADVIA Centaur and ADVIA Centaur XP analyzers.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1657-2014