BD Kiestra ReadA lab automation system recalled for connectivity issues after reboot
BD Kiestra ReadA lab automation systems (Catalog No. 446948) may occasionally lose remote connectivity after a system reboot, which could temporarily delay plate retrieval from the incubator. The issue is intermittent and does not affect test results, assay performance, or sample integrity. This affects 82 units distributed across the United States and internationally.
From the FDA notice · Read the official notice ↗
What to do
- If you operate a BD Kiestra ReadA system, contact BD for guidance on your specific unit and any available updates or fixes
- Monitor system connectivity after any reboots and manually verify module status if needed
- Report any connectivity issues to BD immediately
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- BD Kiestra" ReadA; Catalog No.: 446948.
- Sold in
- Domestic: KS, MD, MN, NJ, TX, VA; International: Australia, Austria, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Kuwait, Mexico, Morocco, Netherlands, Norway, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, United Kingdom;
- Made for
- All ages (under 100 years)
Details
- Units affected
- 82 units (7 US, 65 OUS)
- Illnesses reported
- None reported
- Manufactured in
- Marconilaan 6, Drachten, Netherlands
- Classification
- Class II
- Recall number
- Z-1658-2026
In the FDA's words
- Description
- BD Kiestra" ReadA; Catalog No.: 446948. In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity. Catalog No.: 446948; UDI-DI: 00382904469481; Serial Numbers: All; Distribution: Domestic: KS, MD, MN, NJ, TX, VA; International: Australia, Austria, Belgium, Canada, China, Denmark, France, Germany, Hong Kong, Italy, Kuwait, Mexico, Morocco, Netherlands, Norway, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, United Kingdom; Affected quantity: 82 units (7 US, 65 OUS)
- Hazard
- In certain situations, following a system reboot, one or more modules may have become unreachable via the remote connectivity interface. When this occurred, there was potential for delays in plate retrieval from the incubator while connectivity was restored. This issue was intermittent and did not affect test results, performance of the assay, or sample integrity.