FDA-Device2014-06-04Class II

Medtronic Temporary Pacemaker, Model 5348. Single Chamber External Pulse Generator (EPG) Temporary Pacemaker (AAI/VVI). Used in conjunction with a cardiac pacing lead system for temporary atrial or ventricular pacing in a clinical environment. The device is typically connected to temporary transvenous, epicardial or myocardial pacing leads, in a bipolar configuration, using a patient cable (Medtronic Model 5433A or 5433V) or a surgical cable (Medtronic Model 5832 or 5832S).

Medtronic Inc. Cardiac Rhythm Disease Management
Hazard

Events related to a pacing rate outside of the intended setting, including events of sudden increased pacing rate up to the maximum setting of 180 pulses per minute (ppm), have been reported. This is caused by the development of high resistance on internal electrical connector contacts due to oxidation over time.

Sold states
Worldwide Distribution -- USA, including the territories of Guam and Puerto Rico; and, the countries of Albania, Algeria, Australia, Austria, Bahamas, Bangladesh, Belarus, Belgium, Benin, Bolivia, Bosnia And Herzegovina, Brazil, Bulgaria, Burkina Faso, Chile, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, El Salvador, Estonia, Faroe Islands, Finland, France, Georgia, Germany, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, India, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macau, Malaysia, Malta, Martinique, Mauritius, Mexico, Morocco, Nepal, Netherlands, Netherlands Antilles, New Caledonia, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Saudi Arabia, Senegal, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, and Yemen.
Affected count
16,337 (8,989 US; 7,348 OUS)
Manufactured in
8200 Coral Sea St NE, Saint Paul, MN, United States
Products
Medtronic Temporary Pacemaker, Model 5348. Single Chamber External Pulse Generator (EPG) Temporary Pacemaker (AAI/VVI). Used in conjunction with a cardiac pacing lead system for temporary atrial or ventricular pacing in a clinical environment. The device is typically connected to temporary transvenous, epicardial or myocardial pacing leads, in a bipolar configuration, using a patient cable (Medtronic Model 5433A or 5433V) or a surgical cable (Medtronic Model 5832 or 5832S).

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1661-2014