FDA-Device2014-06-04Class II
Visitec Anterior Chamber Cannula, .40x22mm (27Gx7/8in) REF 585006 Cannulae are intended to irrigate and or aspirate ophthalmic solutions and liquids in or out of the eye, and remove materials where appropriate. These products are used in all sections of the eye. This cannula is a general use cannula mainly designed for the precise maintenance of optimum anterior chamber depth
Beaver-Visitec International Inc.
Hazard
Contains a sharp tip needle when the product should contain a blunt tip cannula
Sold states
Nationwide Distribution including AL, CT, IN, NC, NY, TX, and VA.
Affected count
9970 units
Manufactured in
411 Waverley Oaks Rd Ste 229, Waltham, MA, United States
Products
Visitec Anterior Chamber Cannula, .40x22mm (27Gx7/8in) REF 585006 Cannulae are intended to irrigate and or aspirate ophthalmic solutions and liquids in or out of the eye, and remove materials where appropriate. These products are used in all sections of the eye. This cannula is a general use cannula mainly designed for the precise maintenance of optimum anterior chamber depth
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1664-2014Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief