FDA-Device2026-04-08Class IIPROCESSING DEFECT
Philips ALLURA Xper FD10 OR Table recalled for X-ray foot switch malfunction
What this means
Philips has identified a problem with the wired foot switch on certain ALLURA Xper FD10 operating room tables where X-ray imaging may not start or may work only intermittently. Two units worldwide are affected. No injuries have been reported.
- Check if your facility operates an ALLURA Xper FD10 with System Code 722022 or 722033
- Contact Philips Medical Systems directly for instructions on inspection or repair of the foot switch
- Do not rely on the foot switch until you receive guidance from Philips
The agency's original wording
Hazard
Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
Intended for
Adults
Illnesses reported
None reported
Sold states
Domestic: Nationwide Distribution; International: Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, FrenchPolynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Laos, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkiye, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen;
Affected count
2 units (1 US, 1 OUS)
Manufactured in
Veenpluis 6, Best, Netherlands
Products
ALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033;