Philips Azurion 3 M12 X-ray system recalled for intermittent foot switch failure
Philips identified a problem with the wired foot switch on certain Azurion 3 M12 X-ray imaging systems. The foot switch may fail to start X-ray imaging or cause imaging to work intermittently. About 298 systems worldwide are affected, including 27 in the U.S.
From the FDA notice · Read the official notice ↗ · All Philips recalls
What to do
- Check if your system serial number matches the affected list on the FDA website
- Contact Philips Medical Systems for a replacement foot switch or repair instructions
- Do not use the affected foot switch until you receive guidance from Philips
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229;
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 22 products — see all of them.
Details
- Units affected
- 298 units (27 US, 271 OUS)
- Illnesses reported
- None reported
- Manufactured in
- Veenpluis 6, Best, Netherlands
- Classification
- Class II
- Recall number
- Z-1673-2026
In the FDA's words
- Description
- Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229; Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch. (1) System Code: 722063; UDI: 00884838085275; System Serial Number:122, 56, 64, 131, 84, 132, 96, 98, 108, 77, 82, 145, 47, 146, 80, 102, 43, 117, 2, 149, 99, 44, 3, 28, 144, 62, 119, 61, 22, 29, 153, 40, 70, 30, 25, 19, 120, 121, 110, 14, 16, 17, 53, 103, 127, 97, 143, 41, 104, 116, 151, 93, 141, 115, 137, 73, 150, 128, 105, 4, 50, 129, 72, 38, 21, 78, 23, 76, 63, 69, 59, 51, 92, 139, 130, 154, 75, 24, 94, 100, 54, 46, 26, 71, 106, 83, 36, 37, 27, 15, 135, 123, 107, 87, 85, 89, 88, 86, 90, 91, 79, 133, 81, 140, 52, 48, 148, 55, 113; (2) System Code: 722221; UDI: 00884838085275; System Serial Number:181, 3, 7, 178, 54, 6, 93, 69, 9, 115, 177, 85, 19, 49, 171, 153, 108, 193, 104, 169, 45, 101, 135, 106, 51, 114, 36, 124, 197, 129, 4, 154, 146, 68, 195, 130, 183, 187, 42, 172, 164, 190, 151, 71, 152, 184, 147, 32, 105, 182, 58, 103, 149, 127, 128, 86, 122, 60, 155, 66, 33, 55, 70, 123, 118, 61, 44, 185, 74, 95, 205, 13, 99, 34, 142, 39, 144, 82, 145, 59, 77, 180, 126, 97, 132, 79, 91, 161, 102, 52, 120, 162, 64, 75, 41, 84, 131, 94, 96, 38, 48, 31, 92, 78, 186, 133, 167, 23, 26, 40, 10, 203, 30, 207, 208, 76, 50, 98, 107, 158, 157, 62, 29, 194, 73, 25, 15, 119, 140, 166, 57, 27, 2, 47, 16, 141, 143, 35, 168, 83, 163, 202, 67, 198, 134, 5, 81, 22, 28, 165, 196, 173, 46, 53, 8, 56, 174, 43, 159, 1, 176, 11; (3) System Code: 722229; UDI: 00884838085275; System Serial Number: 22, 9, 40, 7, 24, 35, 16, 15, 41, 5, 3, 29, 4, 38, 33, 32, 6, 2, 20, 26, 31, 10, 11, 21, 25, 13, 8; Distribution: Domestic: Nationwide Distribution; International: Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Fiji, Finland, France, French Polynesia, FrenchPolynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guatemala, Guinea, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Laos, Latvia, Lebanon, Libya, Lithuania, Macao, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkiye, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen; Affected quantity: 298 units (27 US, 271 OUS)
- Hazard
- Philips has identified instances where X-ray imaging may not be initiated or may occur intermittently when using the wired foot switch.
22 products in this recall
Philips X-ray systems recalled for faulty foot switch
Philips is recalling 22 models of its Allura Xper and Azurion X-ray imaging systems. The recall covers a wired foot switch that may fail to start X-ray imaging or may start it intermittently. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- Allura Xper FD10C; System Code: 722001;Z-1659-2026 · 15 units (7 US, 8 OUS)
- ALLURA Xper FD10F; System Code: 722002;Z-1660-2026 · 10 units (8 US, 2 OUS)
- ALLURA Xper FD10; System Code: (1) 722003, (2)722010, (3)722026;Z-1661-2026 · 2291 units (493 US, 1798 OUS)
- ALLURA Xper FD10/10; System Code: (1) 722005, (2)722011, (3)722027;Z-1662-2026 · 112 units (138 US, 422 OUS)
- ALLURA Xper FD20; System Code: (1) 722006, (2)722012, (3)722028;Z-1663-2026 · 5084 units (1616 US, 3468 OUS)
- ALLURA Xper FD20 Biplane; System Code: (1) 722008, (2)722013;Z-1664-2026 · 461 units (198 US, 263 OUS)
- ALLURA Xper FD20 OR Table; System Code: (1) 722015, (2)722023, (3)722035;Z-1665-2026 · 196 units (73 US, 223 OUS)
- ALLURA Xper FD20 Biplane OR Table; System Code: (1) 722020, (2)722025;Z-1666-2026 · 4 units (2 US, 2 OUS)
- ALLURA Xper FD10 OR Table; System Code: (1) 722022, (2)722033;Z-1667-2026 · 2 units (1 US, 1 OUS)
- Allura Xper FD20/10; System Code: 722029;Z-1668-2026 · 85 units (20 US, 65 OUS)
- Allura Xper FD20/20; System Code: 722038;Z-1669-2026 · 111 units (31 US, 80 OUS)
- Allura Xper FD20/20 OR Table; System Code: 722039;Z-1670-2026 · 5 units (OUS only)
- Allura Xper FD20/15; System Code: 722058;Z-1671-2026 · 296 units (68 US, 228 OUS)
- Allura Xper FD20/15 OR Table; System Code: 722059;Z-1672-2026 · 2 units (2 US, 6 OUS)
- Azurion 3 M12; System Code: (1)722063, (2)722221, (3)722229; (this page)Z-1673-2026 · 298 units (27 US, 271 OUS)
- Azurion 3 M15; System Code: (1)722064, (2)722222, (3)722230, (4)722280 (OUS Only);Z-1674-2026 · 798 units (42 US, 756 OUS)
- Azurion 7 B12; System Code: (1)722067, (2)722225, (3)722235;Z-1675-2026 · 670 units (106 US, 564 OUS)
- Azurion 7 B20; System Code: (1)722068, (2)722226, (3)722236;Z-1676-2026 · 1523 units (322 US, 1201 OUS)
- Azurion 7 M12; System Code: (1)722078, (2)722223, (3)722233;Z-1677-2026 · 1994 units (360 US, 1634 OUS)
- Azurion 7 M20; System Code: (1)722079, (2)722224, (3)722234, (4)722282 (OUS only);Z-1678-2026 · 5167 units (1728 US, 3439 OUS)
- Azurion 5 M12; System Code: (1)722227, (2)722231;Z-1679-2026 · 345 units (10 US, 335 OUS)
- Azurion 5 M20; System Code: (1)722228, (2)722232, (3) 722281 (OUS only);Z-1680-2026 · 920 units (168 US, 752 OUS)
More from Philips
See all- Philips Allura Xper systems recalled for FlexVision software issue
ModerateFDANov 6, 2026
- Philips Azurion 3 M12 systems recalled over unapproved drives
ModerateFDAOct 27, 2026
- Philips Azurion X-ray systems recalled over image and movement software issues
ModerateFDAAug 19, 2026
- Philips Azurion X-ray systems recalled for software image and movement issues
ModerateFDAAug 19, 2026
- Philips Allura Xper FD X-ray systems recalled over firmware issue
ModerateFDAAug 19, 2026