FDA-Device2021-06-02Class II
Centaur Syphilis Assay-SYP Kit, 200 Test - Product Usage: For in vitro diagnostic use in the qualitative determination of antibodies to Treponema pallidum in human serum or plasma using the ADVIA Centaur XP/XPT System and ADVIA Centaur CP Systems, SMN:10492493
Siemens Healthcare Diagnostics, Inc
Hazard
Potential for the ADVIA Centaur Syphilis assay to carryover into other commercial assays. This carryover effect only occurs when the impacted assays are used immediately following testing with the Syphilis assay on the ADVIA Centaur XP/XPT or ADVIA Centaur CP systems
Sold states
Worldwide distribution - US Nationwide distribution and the countries of Austria, Belgium, Burkina Faso, Croatia, Czech Republic, Egypt, Finland, France, Fren.Polynesia, Georgia, Germany, Germany, Greece, Guadeloupe, Italy, Kazakhstan, Latvia, Netherlands, Norway, Poland, Portugal, Qatar, Russian Fed., Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, U.A.E., Ukraine, United Kingdom.
Affected count
22550 units
Manufactured in
333 Coney St, N/A, East Walpole, MA, United States
Products
Centaur Syphilis Assay-SYP Kit, 200 Test - Product Usage: For in vitro diagnostic use in the qualitative determination of antibodies to Treponema pallidum in human serum or plasma using the ADVIA Centaur XP/XPT System and ADVIA Centaur CP Systems, SMN:10492493
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1674-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.