FDA-Device2017-04-05Class II

JResultNet drivers using ASTM coding language. Calculator/data processing module, for clinical use - Stand-alone software package that provides connectivity between in-vitro diagnostic instruments and laboratory information's systems, aiding the flow of laboratory information, including test orders and results, specimen information, and patient information

Data Innovations, LLC
Hazard

Software issue resulting in the results of one patient potentially being associated with another.

Sold states
Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, & DC.
Affected count
657 units
Manufactured in
120 Kimball Ave Ste 100, South Burlington, VT, United States
Products
JResultNet drivers using ASTM coding language. Calculator/data processing module, for clinical use - Stand-alone software package that provides connectivity between in-vitro diagnostic instruments and laboratory information's systems, aiding the flow of laboratory information, including test orders and results, specimen information, and patient information

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1680-2017

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief