FDA-Device2021-06-02Class II

VITROS CoV2 Ag calibrator (Product Code 6199950) - Product Usage: is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) and anterior nasal swab specimens collected in CDCs formulation of VTM.

Ortho-Clinical Diagnostics, INc.
Hazard

Ortho investigated and confirmed the potential for unexpected failed calibration, and condition codes associated with the calibration, when using VITROS SARS-CoV-2 Ag Reagent Pack along with VITROS Immunodiagnostic Products SARS-CoV-2 Ag Calibrator, Lot 20.

Sold states
Worldwide distribution - US Nationwide distribution in the states of AR, CA, IA, IL, LA, MO, MS, TX, WI, and the countries of Colombia, India, Japan, Italy, Spain, Portugal, France.
Affected count
74 units
Manufactured in
513 Technology Blvd, N/A, Rochester, NY, United States
Products
VITROS CoV2 Ag calibrator (Product Code 6199950) - Product Usage: is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) and anterior nasal swab specimens collected in CDCs formulation of VTM.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1688-2021

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief