FDA-Device2019-06-12Class II

CyberKnife VSI, Part Numbers: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan b) 033000-010: CyberKnife VSI Robotic Radiosurgery System

Accuray Incorporated
Hazard

There is an unintended drop of a secondary collimator housing that can occur during housing Xchange(TM) workflows. Specifically, during an automated housing Xchange, an interlock may interrupt this operation causing the collimator housing to partially attach.

Sold states
US & Italy, United Kingdom, Switzerland, Belgium, Canada, Germany, France, Japan, Poland, Netherlands, Luxembourg, Australia, Qatar, Latvia, Portugal, India, Hungary, Turkey, United Arab Emirates, Thailand
Affected count
43 units
Manufactured in
1240 Deming Way, Madison, WI, United States
Products
CyberKnife VSI, Part Numbers: a) 032500-010: CyberKnife VSI 9.6.x Robotic Radiosurgery System for Japan b) 033000-010: CyberKnife VSI Robotic Radiosurgery System

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1697-2019