FDA-Device2020-04-22Class II

The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained blood smears from SP-50, selects only smears that require analysis, and transports them to the DI-60 for further analysis. After completing the analysis on DI-60 the smears are placed back in magazines.

Sysmex America, Inc.
Hazard

Software mismatch-When the software versions between the SP-50 and CF-70 are not matched, and an error condition occurs requiring the operator to open the cover door of the CF-70 to remove slides or slide magazines from the CF-70, the operation of the CF-70 does not halt.

Sold states
USA nationwide distribution in the states of AK, AR, AZ, CA, CO, CT, DE, FL, IL, KY, LA, MA, MD, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, TN, TZ, UT and WA..
Affected count
93 units.
Manufactured in
577 Aptakisic Rd, Lincolnshire, IL, United States
Products
The CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stained blood smears from SP-50, selects only smears that require analysis, and transports them to the DI-60 for further analysis. After completing the analysis on DI-60 the smears are placed back in magazines.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1703-2020