Serious riskFDA DeviceClass II

Medline Airway Exam Kit recalled for rubber fragments in Olympus biopsy valves

Medline's Airway Exam Kit contains Olympus biopsy valves that may shed rubber fragments. These fragments can lodge in a patient's airway and may require removal. This could cause inflammation, low oxygen levels, or extend the procedure time.

From the FDA notice · Read the official notice ↗ · All Medline recalls

What to do

  1. Stop using Airway Exam Kit model DYKE1796 immediately, especially lots 24CBT023, 24FBB701, 24LBL343, and 25GBP141.
  2. Contact Medline or your device distributor for instructions on returning or disposing of affected kits.
  3. If a kit was already used on a patient, notify the patient's healthcare provider so they can monitor for symptoms or complications.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Airway Exam Kit, DYKE1796
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 7 products — see all of them.

Details

Units affected
14,379 kits total
Illnesses reported
None reported
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Classification
Class II
Recall number
Z-1705-2026

In the FDA's words

Description
Airway Exam Kit, DYKE1796 Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure. UDI-DI 10193489874716 (ea) 40193489874717 (case) Lots 24CBT023 24FBB701 24LBL343 25GBP141 Distribution: US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX. Affected quantity: 14,379 kits total
Hazard
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

7 products in this recall

Medline kits with Olympus biopsy valves recalled for fragment risk

Medline is recalling 7 kit products that contain Olympus single-use biopsy valves, totaling about 14,379 kits. Olympus found more complaints of rubber fragments detaching from a slit in the valves. A loose fragment could lodge in a patient's airway and may need to be removed in a procedure. No injuries have been reported.

The agency lists each product separately. Check every one you might have.