Moderate riskFDA DeviceClass II

Medline Anterior Hip Pack recalled for Olympus biopsy valve rubber fragments

Medline's Anterior Hip Pack surgical kits contain Olympus biopsy valves that may shed rubber fragments. These fragments could lodge in a patient's airway during a procedure, potentially requiring removal and causing breathing problems or inflammation.

From the FDA notice · Read the official notice ↗ · All Medline recalls

What to do

  1. Stop using Medline Anterior Hip Pack DYNJ64672B (Lot 23LMD300) immediately
  2. Check your facility's inventory for affected kits using UDI-DI 10195327377496 or 40195327377497
  3. Contact Medline Industries or the FDA for instructions on returning or destroying unused kits
  4. Review patient records if kits were already used and report any airway complications to your healthcare provider

Is yours affected?

Compare your label or packaging with the recalled product.

Product
MEDLINE ANTERIOR HIP PACK DYNJ64672B
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 7 products — see all of them.

Details

Units affected
14,379 kits total
Illnesses reported
None reported
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Classification
Class II
Recall number
Z-1711-2026

In the FDA's words

Description
MEDLINE ANTERIOR HIP PACK DYNJ64672B Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure. UDI-DI 10195327377496 (EA) 40195327377497 (CS) LOT 23LMD300 Distribution: US Nationwide distribution in the states of CA, FL, IL, IN, KY, MA, MD, MT, NY, OH, PA, TN, TX. Affected quantity: 14,379 kits total
Hazard
Kits contain recalled Olympus biopsy valves. Olympus is investigating an increase in complaints associated with rubber fragment detachment in the slit of single use biopsy valves. Potential consequences associated with detachment of fragments from valve include foreign body in patient's tracheobronchial tree, which may require intervention for removal. Other potential consequences include inflammatory response, hypoxia, and prolongation of procedure.

7 products in this recall

Medline kits with Olympus biopsy valves recalled for fragment risk

Medline is recalling 7 kit products that contain Olympus single-use biopsy valves, totaling about 14,379 kits. Olympus found more complaints of rubber fragments detaching from a slit in the valves. A loose fragment could lodge in a patient's airway and may need to be removed in a procedure. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

More from Medline

See all

Related topics

OtherParticulateMedical devicesSold nationwideMore from FDA Device