FDA-Device2015-06-17Class II

EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.

Advanced Sterilization Products
Hazard

Advanced Sterilization Products (ASP) is recalling the EVOTECH Endoscope Cleaner and Reprocessor (ECR) because it may not detect an improper endoscope connection on two channels, and subsequently pass the cycle without cancellation if there is a disconnected scope.

Sold states
Worldwide Distribution-US (nationwide) and the countries of Australia, Singapore, Canada, and Thailand.
Affected count
825 units
Manufactured in
33 Technology Dr, N/A, Irvine, CA, United States
Products
EVOTECH Endoscope Cleaner and Reprocessing System 208V, P/N 50004. The ASP EVOTECH ECR Endoscope Cleaner and Reprocessor is a washer/disinfector, which is indicated for use with high-level disinfectant CIDEX OPA Concentrate and an enzymatic detergent (CIDEZYME Gl) to achieve cleaning and high level disinfection of heat sensitive (>60 ¿C) semi-critical endoscopes.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1714-2015

Don't want to check this manually?