FDA-Device2016-06-01Class II

Accriva- APTT Cuvette for use on the Hemochron Jr. test system, catalog #J103, lot B6JCA012.

Accriva Diagnostics Inc., dba ITC, dba Accumetrics
Hazard

Clot activating reagent mis-positioned or visibly absent in the cuvette's test channel causing APTT test to either have an error code or an erroneously high result.

Sold states
U.S. distribution to the following; state: VA. Foreign distribution to the following; China, Italy, Germany.
Affected count
135 boxes
Manufactured in
6260 Sequence Dr, N/A, San Diego, CA, United States
Products
Accriva- APTT Cuvette for use on the Hemochron Jr. test system, catalog #J103, lot B6JCA012.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1720-2016

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief