FDA-Device2018-05-16Class II
BEQ-T 25501 3/8" Better Bladder Pack, Custom Tubing Kit, Catalog No. 709000370
Datascope Corporation
Hazard
The Better Bladder device provided as part of some custom tubing kits has been recalled by its manufacturer. Seal separation from the housing may allow air to enter the housing and collapse the balloon. This may increase resistance to flow in the venous line and cause a drop in ECMO circuit blood flow.
Sold states
Distributed to accounts in AL, AR, CA, DC, FL, IL, LA, MO, NC, PA, SC, TX, and WA.
Affected count
10
Manufactured in
15 Law Dr, N/A, Fairfield, NJ, United States
Products
BEQ-T 25501 3/8" Better Bladder Pack, Custom Tubing Kit, Catalog No. 709000370
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1723-2018Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief