FDA-Device2014-06-18Class II

Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes a supply voltage of 230 V. Article Number 10446782 utilizes a supply voltage of 110 V. Intended to be used for improving the visibility of objects through magnification and illumination.

Leica Microsystems, Inc.
Hazard

The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).

Sold states
Worldwide Distribution -- USA, including the states of AZ, IL, MA, MO, NC, OR, TX, VA, and WI; and, the countries of Australia, Austria, Chile, China, Colombia, Czech Republic, Greece, Israel, Italy, Japan, Malaysia, Netherlands, Norway, Oman, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, and United Kingdom.
Affected count
103 total
Manufactured in
1700 Leider Ln, N/A, Buffalo Grove, IL, United States
Products
Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes a supply voltage of 230 V. Article Number 10446782 utilizes a supply voltage of 110 V. Intended to be used for improving the visibility of objects through magnification and illumination.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1725-2014

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.