FDA-Device2014-06-18Class II
Alere Triage¿ D-Dimer Test PN 98100, Lot W53884B. The Alere Triage¿ D-Dimer Test is a fluorescence immunoassay to be used with the Alere Triage¿ Meters for the quantitative determination of cross-linked fibrin degradation products containing D-dimer in EDTA anticoagulated whole blood and plasma specimens. . The Alere Triage¿ D-Dimer test is intended for use as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic even
Alere San Diego, Inc.
Hazard
Alere initiated this recall because a limited number of Alere Triage¿ D-Dimer devices from PN 98100 Lot W53884B were incorrectly manufactured. As a result, when testing patient sample these affected devices will either generate an error code (e.g. E4 error), or they may generate an incorrect result.
Sold states
Nationwide Distribution only.
Affected count
373 kits
Manufactured in
9975 Summers Ridge Rd, N/A, San Diego, CA, United States
Products
Alere Triage¿ D-Dimer Test PN 98100, Lot W53884B. The Alere Triage¿ D-Dimer Test is a fluorescence immunoassay to be used with the Alere Triage¿ Meters for the quantitative determination of cross-linked fibrin degradation products containing D-dimer in EDTA anticoagulated whole blood and plasma specimens. . The Alere Triage¿ D-Dimer test is intended for use as an aid in the assessment and evaluation of patients suspected of having disseminated intravascular coagulation or thromboembolic events including pulmonary embolism.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1727-2014Don't want to check this manually?