FDA-Device2014-06-18Class II

HLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with water is approximately 450 pounds. The dimensions of the system are: Width: 26.5" (67 cm); Depth: 33" (83 cm); Height: 46.5" (118 cm) to top of control panel; 64" (162.6 cm)- height clearance with lid open. Product Usage: Usage: This device washes reusable medical devices and disinfects them with high temperature hot water (pasteurization). The devices are washed with a cleaning soluti

Cenorin, LLC
Hazard

Cenorin LLC is voluntarily recalling HLD Systems Model 610 Washer/High Level Disinfector due to a potential risk of the system not injecting the proper amount of cleaning fluid during the wash cycle and not achieving the desired level of debris reduction.

Sold states
Distribution USA Nationwide
Affected count
25 HLD systems
Manufactured in
6324 S 199th Pl Ste 107, N/A, Kent, WA, United States
Products
HLD Systems Model 610 Washer/High Level Disinfector. Maximum weight of the Model 610 when filled with water is approximately 450 pounds. The dimensions of the system are: Width: 26.5" (67 cm); Depth: 33" (83 cm); Height: 46.5" (118 cm) to top of control panel; 64" (162.6 cm)- height clearance with lid open. Product Usage: Usage: This device washes reusable medical devices and disinfects them with high temperature hot water (pasteurization). The devices are washed with a cleaning solution (containing NaClO (sodium hypochlorite), surfactants, and detergents) to aid in the removal of debris and visible stains. Next, the medical devices and rinsed two times with water. The final, disinfection step uses only hot water at 160 F ¿ 2 F for 30 minutes.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1729-2014