FDA-Device2024-05-08Class II
ZPLP Distal Lateral Fibular Plate, Left, 10 Holes, 158 MM, Nonsterile-Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-018-10
Zimmer, Inc.
Hazard
Thread form issue for the locking holes in that the locking screws may not properly mate with the plate. The improperly mated screw may not be readily recognizable by the user since the screw may not correctly lock, may result in delay in surgery or loss of fixation resulting further intervention
Sold states
Worldwide distribution - US Nationwide and the countries of Brazil, Canada, China, Hong Kong, India, Japan, New Zealand, Seoul Korea, Singapore, Switzerland, Taiwan.
Manufactured in
1800 W Center St, Warsaw, IN, United States
Products
ZPLP Distal Lateral Fibular Plate, Left, 10 Holes, 158 MM, Nonsterile-Indicated for temporary internal fixation and stabilization of osteotomies and fractures Item Number: 00-2357-018-10
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1729-2024Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief