R3(TM) 22 MM I.D., 52 MM O.D. CONSTRAINED LINER, REF 71339152, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular

Smith & Nephew, Inc.
Hazard

The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.

Sold states
Worldwide Distribution - US Nationwide in the states of CA, FL, IL, IN, MI, NE, NY, TN, TX, VA, WI and the countries of Australia, Switzerland, Chile, Canada, and Spain
Affected count
17 units
Manufactured in
1450 E Brooks Rd, N/A, Memphis, TN, United States
Products
R3(TM) 22 MM I.D., 52 MM O.D. CONSTRAINED LINER, REF 71339152, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained liner is intended for single use only.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1735-2015

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief