FDA-Device2015-06-17Class II
R3(TM) 22 MM I.D., 56 MM O.D. CONSTRAINED LINER, REF 71339156, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular
Smith & Nephew, Inc.
Hazard
The R3 Constrained outer liner locking groove diameter is 0.030 inches oversized and may not allow the locking ring to engage into the R3 shell.
Sold states
Worldwide Distribution - US Nationwide in the states of CA, FL, IL, IN, MI, NE, NY, TN, TX, VA, WI and the countries of Australia, Switzerland, Chile, Canada, and Spain
Affected count
0
Manufactured in
1450 E Brooks Rd, N/A, Memphis, TN, United States
Products
R3(TM) 22 MM I.D., 56 MM O.D. CONSTRAINED LINER, REF 71339156, QTY: (1), STERILE EO. Product Usage: The R3 Constrained liner Acetabular System is a cemented or uncemented prosthesis intended to replace a hip joint. The Constrained liner is intended for primary or revision patients at high risk for hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability and for whom all other options to constrained acetabular components have been considered. The R3 Constrained liner is intended for single use only.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1736-2015Don't want to check this manually?