FDA-Device2022-09-21Class II

ZAP-X Radiosurgery System

Zap Surgical Systems
Hazard

Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

Sold states
Worldwide distribution - US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.
Affected count
8
Manufactured in
590 Taylor Way, San Carlos, CA, United States
Products
ZAP-X Radiosurgery System

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1742-2022