FDA-Device2020-04-29Class II

Fentanyl Reagent Kit, Product N. CC444, Contains: FENT R1 Reagent and FENT R2 Reagent

Carolina Liquid Chemistries Corp
Hazard

Labeled for diagnostic use and sold to clinical labs without a 510(k). Firm corrected labeling to remove clinical diagnostic language and notified customers that products are labeled for "Forensic Use Only" and are not FDA Cleared and not to be used for clinical purposes

Sold states
U.S.: NC, TX, FL, NY, AZ, NJ, NH, CA
Affected count
90 test kits/devices
Manufactured in
510 W Central Ave Ste C, N/A, Brea, CA, United States
Products
Fentanyl Reagent Kit, Product N. CC444, Contains: FENT R1 Reagent and FENT R2 Reagent

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1743-2020