Medtronic MiniMed 640G pumps recalled for insulin delivery risk
Medtronic MiniMed insulin pumps (Paradigm, 600, and 700 series) can deliver too much or too little insulin when the pump height changes relative to the infusion site. When elevated, excess insulin may cause dangerously low blood sugar. When lowered, insufficient insulin may cause dangerously high blood sugar. This affects about 116,000 pumps worldwide.
From the FDA notice · Read the official notice ↗ · All Medtronic recalls
What to do
- Keep your pump at the same height as your infusion site—do not wear it significantly above or below where the catheter enters your skin
- Contact Medtronic immediately for guidance on safe pump positioning and potential replacement options
- Monitor your blood sugar more frequently until you receive instructions from your pump manufacturer
- Discuss this recall with your doctor or diabetes care team
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752)
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
This recall covers 14 products — see all of them.
Details
- Units affected
- 116,645
- Illnesses reported
- None reported
- Manufactured in
- 18000 Devonshire St, Northridge, CA, United States
- Classification
- Class II
- Recall number
- Z-1743-2026
In the FDA's words
- Description
- MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. Pump: UDI-DI(GTIN): User Guide Name(Part-Version): MiniMed 640G Insulin Pump: 643169999763, 763000002763, 763000002671, 643169999770, 763000002886, 643169729353, 763000002657, 763000002688, 763000002664, 763000002749, 763000002817, 763000002787, 763000002800, 763000002756, 763000002848, 763000002794, 763000002695, 763000002725, 763000002701, 763000002633, 763000002862, 643169729360, 763000002879, 763000002824, 763000002732, 763000002640, 763000002770, 763000002855, 763000002718, 763000002831, 763000065195, 763000065201, 763000083441, 763000015510, 763000015541, 763000015480, 763000015589, 763000015497, 763000068998, 763000015381, 763000015435, 763000015336, 763000015367, 763000069001, 763000015374, 763000015398, 763000015350, 763000192228, 763000015572, 763000015343, 763000015411, 763000015428, 763000015404, 763000015442, 763000015459, 763000015534, 763000015503, 763000194383, 763000194222, 763000194444, 763000194390, 763000194239, 763000198596, 763000194406, 763000194369, 763000015596, 763000194277, 763000194666, 763000194352, 763000198602, 763000192211, 763000194451, 763000015565, 763000194246, 763000194338, 763000194307, 763000194475, 763000194314, 763000155346, 763000194468, 763000194437, 763000194376, 763000194345, 763000194253, 763000194260, 763000194413, 763000194420, 763000192136, 763000192143, 763000165246, 763000165307, 763000165130, 643169629813, 763000194321, 763000194291, 763000165147, 763000165277, 763000165222, 763000165116, 763000165123, 763000165185, 763000165154, 763000165314, 763000165338, 763000165260, 763000165093, 763000165284, 763000165215, 763000165109, 763000165345, 763000258337, 763000258344, 763000317140, 763000319274, 763000319540, 763000319618, 763000317089, 763000319489, 763000319533, 763000319380, 763000319366, 763000319267, 763000319298, 763000319304, 763000319236, 763000319328, 763000319571, 763000319229, 763000319281, 763000319311, 763000319588, 763000319519, 763000319243, 763000347536, 763000319427, 763000404567, 763000319557, 763000319564, 763000319410, 763000319526, 763000319434, 763000319595, 763000319373, 763000319397, 763000402686, 763000319441, 763000319359, 763000319496, 763000319458, 763000404550, 763000402693, 763000317188, 763000317102, 763000317119, 763000319342, 763000402709, 763000317195, 763000367039, 763000367046, 763000317157, 643169554931, 643169554948, 643169554917, 643169629691, 643169554955, 643169629677, 643169629653, 643169629684, 643169554924, 643169629660, 643169596429, 643169521049, 643169596405, 643169521315, 643169522138, 643169577664, 643169577688, 643169577695, 643169596368, 643169596382, 643169521865, 643169596443, 643169520882, 643169662544, 643169672239, 643169672246, 643169672260, 643169672277, 643169672253, 643169662582, 643169682672, 643169662551, 643169662568, 643169681972, 643169662575, 643169683211, 643169682405, 643169682139, 643169682788, 643169682245, 643169739468, 643169739451, 643169739482, 643169739444, 643169739512, 643169739475, 643169682948, 643169999800, 643169721289, 643169668874, 643169999787, 763000115968, 763000013066, 763000074814, 763000074821, 763000192204, 763000192198, 763000253288, 763000324704, 763000324698, 763000324681, 00763000115968: MINIMED 640G SYSTEM USER GUIDE(MP6025958-116 1, MP6025958-393 1, M052248C002 1, M052249C002 1). Distribution: Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech R
- Hazard
- All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
14 products in this recall
Medtronic MiniMed insulin pumps recalled for dosing error risk
Medtronic is recalling 14 MiniMed insulin pump models, including the Paradigm, 600 series, and BLE 700 series. Changes in pump height relative to the infusion site can cause the pump to deliver too much or too little insulin. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- MiniMed 670G Insulin Pump (MMT-1760, MMT-1761, MMT-1762, MMT-1780, MMT-1781, MMT-1782)Z-1739-2026 · 84,248
- MiniMed 630G Insulin Pump (MMT-1714, MMT-1715, MMT-1754, MMT-1755)Z-1740-2026 · 60,880
- MiniMed 780G Insulin Pump (MMT-1884, MMT-1885, MMT-1886, MMT-1894, MMT-1895, MMT-1896)Z-1741-2026 · 453,144
- MiniMed 770G Insulin Pump (MMT-1880, MMT-1881, MMT-1882, MMT-1890, MMT-1891, MMT-1892)Z-1742-2026 · 95,704
- MiniMed 640G Insulin Pump (MMT-1711, MMT-1712, MMT-1751, MMT-1752) (this page)Z-1743-2026 · 116,645
- MiniMed 700G Insulin Pump (MMT-1801, MMT-1805, MMT-1850, MMT-1851)Z-1744-2026 · 13,811
- MiniMed 720G Insulin Pump (MMT-1809, MMT-1810, MMT-1859, MMT-1860, MMT-1867)Z-1745-2026 · 37,458
- MiniMed 740G Insulin Pump (MMT-1811, MMT-1812, MMT-1861, MMT-1862)Z-1746-2026 · 24,946
- MiniMed 620G Insulin Pump (MMT-1750)Z-1747-2026 · 799
- Paradigm Insulin Pump (MMT-712, MMT-715)Z-1748-2026 · 73,656
- Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)Z-1749-2026 · 29,074
- Paradigm REAL-Time Revel Insulin Pump (MMT-523, MMT-723)Z-1750-2026 · 26,647
- MiniMed 530G Insulin Pump (MMT-551, MMT-751)Z-1751-2026 · 13,429
- Paradigm REAL-Time Veo Insulin Pump (MMT-554, MMT-754)Z-1752-2026 · 55,724
More from Medtronic
See all- Medtronic MiniMed Go app recalled for failed long-acting dose alerts
ModerateFDAOct 22, 2026
- Medtronic MiniMed CareLink Clinic software recalled for insulin dosing calculation error
SeriousFDAOct 21, 2026
- Medtronic MiniMed Flex insulin pumps recalled for possible malfunction
SeriousFDAOct 14, 2026
- MiniMed 640G loaner pumps recalled for possible cleaning damage
ModerateFDAAug 26, 2026