Medtronic MiniMed Paradigm insulin pumps recalled for incorrect insulin delivery based on pump position
Stop using Medtronic MiniMed Paradigm insulin pump
Medtronic MiniMed insulin pumps (Paradigm, 600, and 700 series) can deliver too much or too little insulin when the pump is held at different heights relative to the infusion site. Too much insulin can cause dangerously low blood sugar; too little can cause dangerously high blood sugar and serious complications.
- Contact Medtronic immediately for instructions on your specific pump model
- Do not rely on your pump until you receive guidance from Medtronic
- Work with your healthcare provider to establish a temporary insulin management plan
- Keep your pump level with your infusion site when in use, and monitor blood sugar closely
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1749-2026More Medtronic MiniMed, Inc. recalls
- FDA-Device2026-04-08Medtronic MiniMed insulin pumps recalled for incorrect insulin delivery based on pump position
- FDA-Device2026-04-08Medtronic MiniMed insulin pumps recalled for incorrect insulin delivery based on pump position
- FDA-Device2026-04-08Medtronic MiniMed insulin pumps recalled for incorrect insulin delivery based on pump position
- FDA-Device2026-04-08Medtronic MiniMed insulin pumps recalled for incorrect insulin delivery based on pump position
- FDA-Device2026-04-08Medtronic MiniMed insulin pumps recalled for incorrect insulin delivery based on pump position
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief