Medtronic MiniMed Paradigm pumps recalled for insulin delivery errors

Medtronic MiniMed insulin pumps (Paradigm, 600, and 700 series) can deliver too much or too little insulin when the pump is held at different heights relative to the infusion site. Too much insulin can cause dangerously low blood sugar; too little can cause dangerously high blood sugar and serious complications.

From the FDA notice · Read the official notice ↗ · All Medtronic recalls

What to do

  1. Contact Medtronic immediately for instructions on your specific pump model
  2. Do not rely on your pump until you receive guidance from Medtronic
  3. Work with your healthcare provider to establish a temporary insulin management plan
  4. Keep your pump level with your infusion site when in use, and monitor blood sugar closely

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722)
Sold in
Nationwide
Made for
Adults (0 months and older)

This recall covers 14 products — see all of them.

Details

Units affected
29,074
Illnesses reported
None reported
Manufactured in
18000 Devonshire St, Northridge, CA, United States
Classification
Class II
Recall number
Z-1749-2026

In the FDA's words

Description
Paradigm REAL-Time Insulin Pump (MMT-522, MMT-722) All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. Pump: UDI-DI(GTIN): User Guide Name(Part-Version): Paradigm Real-Time Insulin Pump: 00613994409423, 00613994622600, 613994622747, 613994407733, 00613994407740, 00613994136374, 00613994271600, 00613994106476, 00613994373120, 00613994409478, 00613994312426, 00613994151780, 00885074419831, 00613994622747, 00613994951502, 00613994622648, 00643169026100, 00613994622624, 00613994622556, 00613994402011, 00613994409553, 00613994368485, 00613994368508, 00613994368522, 00613994108739, 00613994368607, 00613994368492, 00613994368515, 00613994123084, 00613994124814, 00613994402523, 00613994402530, 00613994402516, 00613994401502, 00613994402509, 00613994405913, 00613994123206, 00643169688421, 00643169199699, 00643169199675, 00613994409461, 00613994124777, 00643169572492, 00613994267177, 00643169848054, 00613994266897, 00613994310804, 00613994310798, 00613994310774, 00613994310903, 00613994401564, 00613994401663, 00613994402554, 00613994402462, 00613994402592, 00613994402608, 00613994405944, 00613994402455, 00613994402547, 00613994402479, 00763000069292, 00613994402578, 00613994409447, 00643169729438, 00763000069254, 00763000084325, 00613994402486, 00613994405920, 00763000069261, 00613994948595, 00613994929778, 00613994224569, 00613994929808, 00613994930361, 00613994929761, 00643169199705, 00643169199668, 00643169496248, 00643169848016, 00643169848023, 00763000161095,00613994123329, 00613994125026, 00613994247728, 00613994124692, 00613994124739, 00613994266903, 00613994267153, 00613994266880, 00613994267207, 00613994310866, 00613994310811, 00613994310781, 00613994310873, 00613994310613, 00613994310910, 00613994310880, 00613994310767, 00613994375025, 00613994368638, 00613994368553, 00613994368560, 00613994405937, 00613994368539, 00613994368546, 00613994377586, 00613994368591, 00613994368584, 00613994368577, 00613994402585, 00613994402561, 00613994401588, 00613994929815, 00643169221970, 00763000069247, 00763000090654, 00763000640163, 00613994601629, 00643169354975, 00613994401977, 00763000084332, 00643169919389, 00763000161118, 00763000161071: Artwork, CRB D Process, Instr, 522/722, NA(MP6025250-013DOC); Artwork, CRB D Process, Paradigm¿ 522 and 722 Insulin Pumps User(MP6025681-022DOC); Artwork, CRB A Process, Instr, 522/722,Block Mode,TST,Nordic,EN(MP6025681-023DOC); Artwork, CRB D Process, Instr, 522/722 with Enlite, OUS, EN(MP6025646-021DOC); Artwork, CRB D Process, Instr, 522/722, w/Enlite, OUS, EN(MP6025646-022DOC); Artwork, CRB A Process, Instr, 522/722, w/Enlite, OUS, EN(MP6025646-023DOC); Artwork, CRB A Process, Instr, 522/722, w/Enlite, OUS, EN(MP6025646-023DOC); Artwork, New Product, Feature (Form, Fit, or Function), Regulatory, Technical Literature, CRB; IFU, MMT-522/722 PRDGM X22 PUMP W/ENLITE, EN, OUS(MP6025646-024DOC); Artwork, New Product, Feature (Form, Fit, or Function), Regulatory, Technical Literature, CRB; IFU, MMT-522/722 PRDGM X22 PUMP W/ENLITE, EN, OUS(MP6025646-024DOC); Artwork, New Product, Feature (Form, Fit, or Function), Regulatory, Technical Literature, CRB; IFU, MMT-522/722 PRDGM X22 PUMP W/ENLITE, EN, OUS - Nordic cover (MP6025681-024DOC); Distribution: Worldwide - US Nationwide distribution including in the states of OR, CA, AZ, NY,TX,FL, OH, MO, TN,NJ, NC, SC, IL, VA, WA, DC, GA, PA, MS,IA, MI, NH, OK, SD, MT, WI, WV, MA, LA, MN, RI, AR, ME, AL, IN, UT, KY, CO, ND, MD, CT, NM, KS, ID, NV, DE, NE, VT, AK, HI, WY, PR, VI and the countries of Taiwan, New Zealand, Australia, Bangladesh, Brunei Darussalam, Cambodia, Indonesia, Malaysia, Maldives, Philippines, Thailand, Vietnam, Singapore, Pakistan, Hong Kong, China, Argentina, Aruba, Barbados, Bermuda, Bolivia, Plurinational State Of, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, El Salvador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, Puerto Rico, Trinidad And Tobago, Uruguay, Venezuela, Canada, South Korea, India, Albania, Algeria, Andorra, Armenia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bosnia And Herzegovina, Botswana, Bulgaria, Canary Islands, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Faroe Islands, Finland, France, French Guyana, French Polynes., Georgia, Germany, Gibraltar, Greece, Guadeloupe, Hungary, Iceland, Iran, Iraq, Ireland, Isr
Hazard
All Medtronic MiniMed infusion pumps (Paradigm series, 600 series, and BLE 700 series) were found to be affected by unintended over- and under-delivery of insulin when the pump is elevated or lowered relative to the infusion site, respectively, because of changes in gravitational force impacting hydrostatic and hydrodynamic pressures. Insulin over-delivery, which can occur when the pump is elevated relative to the infusion site, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An under-delivery of insulin, which can occur at lowered pump height conditions relative to the infusion site, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.

14 products in this recall

Medtronic MiniMed insulin pumps recalled for dosing error risk

Medtronic is recalling 14 MiniMed insulin pump models, including the Paradigm, 600 series, and BLE 700 series. Changes in pump height relative to the infusion site can cause the pump to deliver too much or too little insulin. No injuries have been reported.

The agency lists each product separately. Check every one you might have.