FDA-Device2018-05-16Class II

Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body

Medtronic Sofamor Danek USA Inc
Hazard

Slot screw had an overall length of 15mm, but per the specification, the screw should have an overall length of approximately 17mm.

Sold states
Worldwide Distribution - US Nationwide in the states of GA, CT, LA, MD, WY, IN, MD, FL, and France
Affected count
39 units
Manufactured in
1800 Pyramid Pl, N/A, Memphis, TN, United States
Products
Medtronic ZEVO(TM) Anterior Cervical Plate System SLOT SCREW, REF 7710015, Size 4.5mm x 15mm Product Appliance, fixation, spinal intervertebral body

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1750-2018