FDA-Device2022-09-21Class II

iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration needle, through the reading chamber, and to the waste on the iSED Automated ESR Analyzer PN: 112-13-002

Alcor Scientific, Inc.
Hazard

Primary Pump Tubing Spare Part is incorrect. If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSEDs Probe Tube

Sold states
US Nationwide Distribution - FL, MA, MD, PA, SD, WI Foreign: Croatia Honduras Korea
Affected count
38 units
Manufactured in
20 Thurber Blvd, Smithfield, RI, United States
Products
iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration needle, through the reading chamber, and to the waste on the iSED Automated ESR Analyzer PN: 112-13-002