FDA-Device2024-05-08Class II
Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.
Johnson & Johnson Surgical Vision, Inc.
Hazard
Due to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.
Sold states
U.S. Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, LA, MA, MD, ME, MN, MO, NC, NJ, NY, OR, PA, SC, TX, VA, WA, and WI.
Affected count
3,428 units
Manufactured in
31 Technology Dr Ste 200, Irvine, CA, United States
Products
Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L The irrigation sleeve is a device intended to direct irrigation solution across the shaft of a phacoemulsification tip, allowing the solution to enter the eye during ocular surgery.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1755-2024Don't want to check this manually?