FDA-Device2024-05-15Class II

PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA¿ TL system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sac

SPINEART SA
Hazard

Due to receiving information regarding difficulties with the insertion of a setscrew inside a screw head. Issues involved, either: 1. The setscrew is not tightened to torque 2. The setscrew is tightened to torque but position askew

Sold states
U.S. Nationwide distribution in the states of CA, and KY. Not provided due to affected devices were manufactured O.U.S.
Affected count
242 systems
Manufactured in
Chemin Du Pre-Fleuri 3, N/A, Plan-Les-Ouates, N/A, Switzerland
Products
PERLA TL 25D SCREW REF: TLF-DS 45 25-S TLF-DS 45 30-S TLF-DS 45 35-S TLF-DS 45 40-S TLF-DS 45 45-S TLF-DS 55 30-S TLF-DS 55 35-S TLF-DS 55 40-S TLF-DS 55 45-S TLF-DS 65 35-S TLF-DS 65 40-S TLF-DS 65 45-S TLF-DS 65 50-S The PERLA¿ TL system is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (painful degeneration of the disc), spondylolisthesis, trauma, spinal stenosis, deformities (i.e. scoliosis, kyphosis, or lordosis), tumor and failed previous fusion (pseudarthrosis).

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1758-2024

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief