FDA-Device2021-06-09Class II
MiniMed 670G Insulin Pump: Models: PUMP MMT-1740KX 670G V6.3 SF MG CLIN;¿ PUMP MMT-1740KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KFX 670G V6.3 SF MG CLIN FR;¿ PUMP MMT-1741KFX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KX 670G V6.3 SF MG CLIN GER;¿ PUMP MMT-1741KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KXS 670G V6.3 SF MG CLIN IS;¿ PUMP MMT-1741KXS 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1742KX 670G V6.1 MM CLIN;¿ PUMP MMT-1742KX 670G V6.3 SF MM CLIN;¿ PUMP MMT-1742KX 670G V6.3 SF MM CLIN SLO;¿ PUMP
Medtronic Minimed
Hazard
Due to a software design issue, under certain conditions, a software fault is detected when a large bolus delivery at quick bolus speed completes. if the user is not aware of the amount of active insulin and delivers an additional bolus, there is a risk of insulin over delivery.
Sold states
U.S.(Clinical): AR, CA, CO, CT, FL, GA, ID, MI, MN, TN, TX, and WA O.U.S.: Austria, Bahrain, Belgium, Canary Islands, Chile, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Gibraltar, Greece, Hungary, Iceland, Ireland, Israel, Italy, Liechtenstein, Luxembourg, Malta, Martinique, Netherlands, New Caledonia, Norway, Poland, Qatar, Reunion, Romania, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, and United Kingdom
Affected count
U.S. Investigational Clinical Study = 1,266 devices
Manufactured in
18000 Devonshire St, Northridge, CA, United States
Products
MiniMed 670G Insulin Pump: Models: PUMP MMT-1740KX 670G V6.3 SF MG CLIN;¿ PUMP MMT-1740KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KFX 670G V6.3 SF MG CLIN FR;¿ PUMP MMT-1741KFX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KX 670G V6.3 SF MG CLIN GER;¿ PUMP MMT-1741KX 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1741KXS 670G V6.3 SF MG CLIN IS;¿ PUMP MMT-1741KXS 670G V6.3 CLIN 020 MG;¿ PUMP MMT-1742KX 670G V6.1 MM CLIN;¿ PUMP MMT-1742KX 670G V6.3 SF MM CLIN;¿ PUMP MMT-1742KX 670G V6.3 SF MM CLIN SLO;¿ PUMP MMT-1742KX 670G V6.3 CLIN 020 MM;¿ PUMP MMT-1742KX 670G V6.3 CLIN 020 EU MM;¿