Moderate riskFDA DeviceClass II

Beckman Coulter MicroScan Neg MIC 3J antibiotic susceptibility panels recalled for manufacturing defects

Beckman Coulter identified a manufacturing defect in certain MicroScan Neg MIC 3J antibiotic susceptibility test panels (Catalog C54814, Lot 2025-08-23). Some panels are missing a drug in one well and contain double the amount of antibiotics in others, which could produce inaccurate test results. No injuries have been reported.

From the FDA notice · Read the official notice ↗ · All Beckman Coulter recalls

What to do

  1. Stop using MicroScan Neg MIC 3J panels with Catalog Number C54814 and Lot Number 2025-08-23
  2. Contact Beckman Coulter for instructions on return or disposal
  3. Review any patient test results from affected panels with your laboratory director

Is yours affected?

Compare your label or packaging with the recalled product.

Product
MicroScan Neg MIC 3J REF C54814
Sold in
Nationwide
Made for
All ages (0 months and older)

Details

Units affected
688 boxes (20 panels per box)
Illnesses reported
None reported
Manufactured in
2040 Enterprise Blvd, West Sacramento, CA, United States
Classification
Class II
Recall number
Z-1772-2026

In the FDA's words

Description
MicroScan Neg MIC 3J REF C54814 Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics Catalog Number: C54814 UDI code: N/A Lot Number: 2025-08-23 Distribution: International distribution in the country of Japan. Affected quantity: 688 boxes (20 panels per box)
Hazard
Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics

More from Beckman Coulter

See all

Related topics